Table 2 Test–retest ICC and SEM of TEOAE measurements between test series

From: Otoacoustic emission measurements: a test–retest reliability study

 

F2 frequency (Hz)

 

1000

1500

2000

3000

4000

AMPLITUDE

ICC

M1–M2

0.94

0.90

0.98

0.99

0.98

M1–M3

0.86

0.85

0.94

0.97

0.97

M2–M3

0.90

0.92

0.95

0.97

0.98

SEM

M1–M2

0.52

0.88

0.15

0.12

0.19

M1–M3

0.92

0.92

0.46

0.32

0.23

M2–M3

0.79

0.67

0.42

0.33

0.22

SNR

ICC

M1–M2

0.89

0.92

0.96

0.97

0.97

M1–M3

0.88

0.91

0.94

0.95

0.95

M2–M3

0.85

0.90

0.95

0.96

0.95

SEM

M1–M2

1.03

0.72

0.33

0.34

0.26

M1–M3

1.13

0.78

0.51

0.45

0.42

M2–M3

1.25

0.83

0.46

0.40

0.41

  1. SEM Standard measurement error, ICC İntraclass correlation coefficient, SNR Signal/noise ratio